Sensocon USA Series DPT1-R8 - Range -1.0 - 01.0 in WC
Sensocon USA Series DPT1-R8 - Range -1.0 - 01.0 in WC

Sensocon USA Series DPT1-R8 - Range -1.0 - 01.0 in WC

Sensocon USA Series DPT1-R8 - Range -1.0 - 01.0 in WC Specification

  • Seal Material
  • Silicone
  • Thread
  • 1/8 NPT Female
  • Outer Size
  • 126 x 90 x 56 mm (approx.)
  • Display
  • 4-digit LCD Digital Display
  • Grade
  • Industrial
  • Sensor Type
  • Differential Pressure Sensor
  • Material
  • ABS Plastic Housing
  • Processing Type
  • Electronic, Microprocessor based
  • Mount Type
  • Wall or Panel Mount
  • Scale Range
  • -1.0 to 1.0 in WC
  • Accuracy
  • 1% of Full Scale
  • Finishing
  • Matte Front with Polycarbonate Casing
  • Measure Pressure Of
  • Air and Compatible Non-corrosive gases
  • Accessories
  • Mounting kit, Calibration Certificate, User Manual
 

Sensocon USA Series DPT1-R8 - Range -1.0 - 01.0 in WC Trade Information

  • Payment Terms
  • Cash on Delivery (COD), Cash Against Delivery (CAD), Cash in Advance (CID), Cheque, Cash Advance (CA)
  • Main Export Market(s)
  • Australia, Middle East, Western Europe, Africa, Central America, South America, Asia, Eastern Europe, North America
  • Main Domestic Market
  • All India
 

About Sensocon USA Series DPT1-R8 - Range -1.0 - 01.0 in WC

Sensocon USA Series DPT1-R8 - Range -1.0 - 01.0 in WC

APPLICATIONS 

DPT-R8 series devices are commonly used in HVAC/R systems for: Fan, blower and filter monitoring Pressure and flow monitoring Valve and damper control Pressure monitoring in clean rooms

WARNING 

READ THESE INSTRUCTIONS CAREFULLY BEFORE ATTEMPTING TO INSTALL, OPERATE OR SERVICE THIS DEVICE.

Failure to observe safety information and comply with instructions could result in PERSONAL INJURY, DEATH AND/ OR PROPERTY DAMAGE.

To avoid electrical shock or damage to equipment, disconnect power before installing or servicing and use only wiring with insulation rated for full device operating voltage.

To avoid potential fire and/or explosion do not use in potentially flammable or explosive atmospheres.

Retain these instructions for future reference.

This product, when installed, will be part of an engineered system whose specifications and performance characteristics are not designed or controlled by Sensocon. Review applications and national and local codes to assure that the installation will be functional and safe. Use only experienced, knowledgeable, technicians to install this device. 

SPECIFICATIONS 

Performance Accuracy (at applied pressure) Ranges < 0.5 inWC = 0.01 inWC Ranges 0.5 inWC = 1.5% inWC Accuracy specifications include: general accuracy, temperature drift, linearity, hysteresis, long term stability, and repetition error

Thermal Effects Temperature compensated across the full spectrum of capability 

 Overpressure Proof Pressure: 100 inWC Burst Pressure: 120 inWC

Zero Point Calibration Automatic with function or Manual push-button Auto-Zero

Response Time 4.0 0.8 sec, Selectable via Jumper

Technical Specifications Media Compatibility Dry air or non-aggressive gases

Measuring Units (select via Jumper) Pa, kPa, mbar, inWC, mmWC Measuring Element Piezoresistive

Environment: Operating Temperature: 14 122F (-10 50C) Storage Temperature: -4 158F (-20 70C) Humidity: 0 to 95 % rH, non condensing 

Physical Case Dimensions: 3.54 x 3.74 x 1.4 (90.0 x 95.0 x 36.0 mm) Weight: 5.3 oz (150 g) Mounting: 2 each 0.1875 (4.3 mm) screw holes, one slotted Materials: Case: ABS (UL 94 V-0 Approved) Lid: PC (UL 94 V-1 Approved) Protection standard: IP54 / NEMA3 

Display (optional) 3 1/2 Digit LCD Backlit Display 2-Line Display (12 character/line) Size: 1.81 W x 0.57 H (46.0 x 14.5 mm) 

Electrical connections: 4 screw terminal block Wire: 12-24AWG (0.2 1.5 mm2 

Cable Entry Strain Relief: M16 Knockout : 0.625 (16mm)

Pressure Fittings 3/16 (5.2mm) barbed brass + High pressure Low pressure 

Electrical Voltage Circuit: 3-wire ([V Out], 24V, GND) Input: 24VAC or VDC, 10% Output: 0-10V or 0-5V, selectable via Jumper Power Consumption: <1.0W Resistance Minimum: 1k

Current Circuit: 3-wire (mA Out, 24V, GND) Input: 24VAC or VDC, 10% Output: 4



Versatile Output & Wide Application

This differential pressure sensor accommodates both 4-20 mA and 0-10 V output signals, easily selectable to suit integration with diverse control systems found in HVAC, cleanrooms, and process control industries. It is engineered for measuring air and compatible non-corrosive gases, offering flexibility in environmental monitoring and safety systems.


Accurate and Reliable Measurements

Ensuring a 1% full scale accuracy, the Series DPT1-R8 delivers dependable performance for sensitive pressure monitoring tasks. With its fast response time (1s) and zero point adjustment accessible from the front panel, it maintains optimal calibration in changing conditions, making it trustworthy for critical applications.


Robust Design for Industrial Environments

Protected by an IP54-rated polycarbonate casing, this sensor withstands challenging industrial settings, while its wall or panel mounting capability expands installation possibilities. The silicone seal and ABS housing ensure durability, and the device arrives complete with necessary mounting accessories and documentation.

FAQs of Sensocon USA Series DPT1-R8 - Range -1.0 - 01.0 in WC:


Q: How do I select between the 4-20 mA and 0-10 V output signals on the Sensocon Series DPT1-R8?

A: You can easily select the desired output signal using the settings accessible via the devices front panel. Refer to the user manual for step-by-step instructions on switching between 4-20 mA and 0-10 V signals to match your system requirements.

Q: What is the recommended installation process for the DPT1-R8 differential pressure sensor?

A: The unit is designed for wall or panel mounting and should be installed using the included mounting kit. Screw terminal electrical connections should be securely made, and the 1/8 NPT female thread used for the pressure ports. Please follow the detailed installation steps provided in the user manual.

Q: When is zero point adjustment necessary, and how is it performed?

A: Zero point adjustment is recommended during initial setup or after prolonged operation to maintain measurement accuracy. To adjust, use the front panel buttons while ensuring no pressure difference is applied across the sensor during the process.

Q: Where can the DPT1-R8 be used given its operating temperature and protection class?

A: This sensor is suited for industrial settings where temperatures range from -10C to +60C. The IP54 protection class enables usage in environments with moderate dust and moisture, but it should be shielded from exposure to corrosive gases or direct water jets.

Q: What gases can the Sensocon DPT1-R8 measure, and what are its limitations?

A: It is specifically designed to measure differential pressure of air and other compatible non-corrosive gases. The device should not be used with corrosive or flammable gases, which could damage its silicone seal and ABS plastic housing.

Q: How can I benefit from the fast response time and accurate readings of the DPT1-R8 in process control?

A: Its 1 second response time and 1% accuracy enhance real-time monitoring, enabling prompt control actions in sensitive environments such as cleanrooms, ensuring proper ventilation and air quality maintenance.

Q: What accessories and documentation are included with the DPT1-R8 pressure sensor?

A: Every unit is supplied with a mounting kit for secure installation, a calibration certificate verifying its accuracy, and a user manual containing comprehensive usage and setup guidance.

Tell us about your requirement
product

Price:

Quantity
Select Unit

  • 50
  • 100
  • 200
  • 250
  • 500
  • 1000+
Additional detail
Mobile number

Email

More Products in Series DPT Differential Pressure Transmitter India Category

Sensocon USA Differential Pressure Transmitter Series DPT1-R8 - Range  0 - 1.25 mbar

Sensocon USA Differential Pressure Transmitter Series DPT1-R8 - Range 0 - 1.25 mbar

Sensor Type : Differential Pressure Transmitter

Finishing : Other, Matte, IP54 enclosure

Outer Size : 96 x 96 x 60 mm

Mount Type : Other, Panel Mount

Grade : Industrial

Display : LED Digital

Sensocon USA Series  DPT1-R8 - Range -0.1 - 0.1 in WC

Sensocon USA Series DPT1-R8 - Range -0.1 - 0.1 in WC

Sensor Type : Differential Pressure Sensor

Finishing : Other, ABS Plastic Housing

Outer Size : 117 x 95 x 40 mm

Mount Type : Other, Wall Mount

Grade : Industrial

Display : LCD Digital

Sensocon USA Differential Pressure Transmitter Series DPT30-R8 - Range  0 - 508 mmWC

Sensocon USA Differential Pressure Transmitter Series DPT30-R8 - Range 0 - 508 mmWC

Sensor Type : Differential Pressure Transmitter

Finishing : Other, ABS Plastic Enclosure

Outer Size : 112 x 81 x 36 mm

Mount Type : Other, Wall Mount

Grade : Industrial Grade

Display : LCD Digital Display

Sensocon USA Differential Pressure Transmitter Series DPT10-R8  - Range   0 - 1.0 inWC

Sensocon USA Differential Pressure Transmitter Series DPT10-R8 - Range 0 - 1.0 inWC

Sensor Type : Differential Pressure Transmitter

Finishing : Other, ABS Plastic Housing

Outer Size : 4.72 x 2.83 x 1.38 (120mm x 72mm x 35mm)

Mount Type : Other, Panel or Surface Mount

Grade : Industrial

Display : No display (Transmitter only)



Back to top