About GEMTECH OMEGA DIFFERENTIAL PRESSURE GAUGE Near MPAKVN New Industrial Area Madhya Pradesh

GEMTECH OMEGA DIFFERENTIAL PRESSURE GAUGE Near MPAKVN New Industrial Area Madhya Pradesh
Series G2000 GEMTECH Differential Pressure Gauges are Widely Utilized For Measuring Pressure of Fan And Blowing Machine, resistance of Filter, Wind Speed,Furnace Pressure, Differential Pressure Of Pore Plate,The Pressure of drum water level,Liquid Amplifier And Hydraulic Pressure System.Meantime,They Are Used To Control The Automatic Valve And Specific Value Of Air-Producer Gas In Combustion And They Are Used To Test Blood Pressure And Respiratory Pressure In Medical Health Care Equipment.
There Are More Than 81 Models To Suit Your Needs Precisely.The Key Components Adopt Import Materi
SPECIFICATION
- Ambient Temperature - -21140-660
- Rated Pressure -20 inHg30 inHg (-68103KPa)
- Overpressure - Protective rubber stopper will be blown out if rated pressure is up to about 50.8 inHg (172KPa)
- Process Connections - 1/8" female NPT duplicate high and low pressure taps-one pair side and one pair back
- Housing - Die cast aluminum case and bezel,With acrylic cover, Exterior finish is coated gray to withstand 168 Hour Salt Spray Corrosion Test
- Accuracy - 2% of full span at 70 (21 (3% on-0,and 4% on -00)
- Range -20 inHg38 inHg (-68KPa130KPa)(Special Specification also can be ordered.
FEATURES
- OEM/ODM
- There Is No Need To Fill Liquid To Gauge
- The Gauge Can Measure Positive Pressure,Negative Pressure And Differential Pressure
- It Will Be Easy To Install And Measure In Different Ways Because Of Two Groups Of Ports Which Can Measure Positive And Negative Pressure
- To Eliminate The Friction From The Movement Of Gear Wheel By The Mechanism Of Magnetic Magnet Feed
- Exact And Stable Fixed Position Of Needle
- The Gauge Has Great Antiknock Characteristics
- There Are 81 Types Of Ranges.
- Minimum is 0-60 Pa And Maximum is 0-21-KPa
PRODUCT APPLICATIONS
- Filter Monitoring
- Air Velocity With Pitot Tube
- Blower Vacuum Monitoring
- Fan Pressure Indication
- Duct,Room or Building Pressures
- Clean Room Positive Pressure Indication
MODELS & ORDERING
GEMTECH By Model Number Popularity
| Model |
Description |
| G2201 |
Differential Pressure Gage,Range 0-1 PSI,Minor Divisions .02 |
| G2202 |
Differential Pressure Gage,Range 0-2 PSI,Minor Divisions .05 |
| G2203 |
Differential Pressure Gage,Range 0-3 PSI,Minor Divisions .10 |
| G2204 |
Differential Pressure Gage,Range 0-4 PSI,Minor Divisions .10 |
| G2205 |
Differential Pressure Gage,Range 0-5 PSI,Minor Divisions .10 |
| G2210 |
Differential Pressure Gage,Range 0-10 PSI,Minor Divisions .20,MP Option Standard |
| G2215 |
Differential Pressure Gage,Range 0-15 PSI,Minor Divisions .50,MP Option Standard |
| G2220 |
Differential Pressure Gage,Range 0-20 PSI,Minor Divisions .50,MP Option Standard |
| G2230 |
Differential Pressure Gage,Range 0-30 PSI,Minor Divisions 1.0,HP Option Standard
|
Additional Information:

Reliable Pressure Monitoring Solution
The GEMTECH Omega Differential Pressure Gauge is engineered for reliability and precision in measuring the pressure of gases across critical industrial processes. With a 4-inch analog dial and metal alloy body, it guarantees durability in environments like power plants and cement factories. This gauge supports a wide range of scale measurements and maintains accuracy even at demanding operational temperatures, making it a dependable choice for professionals.
Versatile Installation & Comprehensive Applications
This gauge boasts duplicate high and low pressure taps (side and back), accommodating diverse installation requirements. It easily fits into pharmaceutical, fertilizer, textile, and steel plant operations, providing consistent performance. The back-mount design and clear dial readability ensure straightforward integration and quick status assessments for plant operators and engineers.
FAQ's of GEMTECH OMEGA DIFFERENTIAL PRESSURE GAUGE Near MPAKVN New Industrial Area Madhya Pradesh:
Q: How is the GEMTECH Omega Differential Pressure Gauge installed?
A: The gauge can be mounted using its back mount configuration. Its 1/8" female NPT duplicate high and low pressure taps (available both on the side and back) make the installation process flexible, allowing users to choose the most suitable connection method for their setup.
Q: What scale ranges are available on the Omega Differential Pressure Gauge?
A: This gauge supports an extensive range of scale units, including MM WC, CM, PAC, KPA, INCH, PSI, and MBAR. Such versatility makes it suitable for a wide range of measurement requirements in various industries.
Q: When is the GEMTECH Omega Differential Pressure Gauge most beneficial to use?
A: This pressure gauge is especially beneficial in critical process control scenarios, such as monitoring clean rooms, ventilation systems, filtration units, and any environment where precise pressure differential measurement for gases is essential.
Q: Where can the gauge be typically used?
A: It is designed for use in pharmaceutical plants, power generation facilities, cement manufacturing, steel mills, fertilizer plants, and textile industries-essentially anywhere gas pressure differentials need accurate and reliable monitoring.
Q: What are the benefits of using this differential pressure gauge?
A: Key benefits include high accuracy (up to 2% of FS), durable metal alloy construction, easy-to-read analog display, and adaptability to various industrial standards and installation needs. These qualities help ensure process integrity and regulatory compliance.
Q: How is measurement accuracy maintained across different pressure ranges?
A: The gauge offers 2% (HA model 1) accuracy of full scale, with specific ranges (like -0, -100PA, -125PA, -10MM) having 3% accuracy, and others (like -00, -60PA, -6MM) at 4% accuracy, ensuring reliable readings throughout its operational range at 70F (21.1C).
Q: What is the process for measuring gas pressure using this gauge?
A: Simply connect the high and low pressure taps to the respective pressure sources. The differential pressure will then be indicated on the clear analog dial, allowing for immediate visual monitoring and easy interpretation by operational staff.