About Galaxy Magnehelic Gauge Model G2000-250MM Range 0-250 MM WC
Galaxy MEGNAHELIC GAUGE Model G2000-250MM Range 0-250 MM WC
The Series G2000 differential pressure gauges are low-cost, diaphragm operated, mechanical pressure gauges,
Series G2000 are for measuring fan and blower pressures, filter resistance, furnace draft, pressure drop across orifice plates, monitoring blood and respiratory pressures in medical care equipment With simple, frictionless movement parts,It can quickly indicate low air or non-corrosive gas pressures, either positive, negative or differential It can resist shock, vibration and over-pressures.
Technical Specification
- Pressure Limits: 20 Fig to 15 PSI (-0.677 bar to 1.034 bar)
- Media Compatibility: Air and compatible gasses
- Accuracy: +1- 2% full scale (+/- 3% an -0 and +/-4% on -00) throughout range at 70F (21 C)
- Temperature Ranges: 20 to 140F (-7 to 60C)
- Housing Material: Die cast aluminum case and bezel (ABS for P2000), with acrylic cover
- Process Connection: 1/8 female NPT duplicate high and low pressure taps one pair side and one pair back.
- Weight: 1 lb 2 oz (510 g)
- Accessories: Two 1/8 NPT barbed frftings,two 1/8 NPT plugs, and three mounting tabs and screws
- Warranty: 1 years Approvals: Roi-IS, CE
Precision and Durability for Industrial NeedsThe Galaxy Magnehelic Gauge is crafted for reliable pressure measurement in various industrial environments. Its rugged die-cast aluminum case, powder-coated for protection, and the external zero adjustment screw offer both durability and easy calibration. Designed for accuracy and consistent performance, it serves as a trustworthy tool for engineers and technicians monitoring air and non-corrosive gases.
Flexible Mounting and Easy IntegrationThis differential pressure gauge is engineered with both side and back process connection options, supporting flexible installation on surfaces or flush mounts. Supplied with mounting brackets and 1/8 NPT plugs, it is versatile for new or retrofit applications. The analog display provides clear readings, making gauge integration straightforward and practical in any compatible system.
FAQ's of Galaxy Magnehelic Gauge Model G2000-250MM Range 0-250 MM WC:
Q: How do I install the Galaxy Magnehelic Gauge Model G2000-250MM?
A: The gauge features surface and flush mount options with both side and back process connections (1/8 NPT female). Use the supplied mounting brackets and plugs for secure installation on compatible panels or surfaces.
Q: What types of media can this gauge measure?
A: This model is specifically designed to measure the differential pressure of air and non-corrosive gases only. It is not suitable for corrosive or liquid media.
Q: When should I use the external zero adjustment screw?
A: The external zero adjustment screw is used to calibrate the pointer to the zero position before measurement or after maintenance, ensuring the accuracy of your readings.
Q: Where is the process connection located on the gauge?
A: Process connections are provided on both the side and back of the gauge, which allows for flexible installation depending on your system's layout and requirements.
Q: What benefits does the Galaxy Magnehelic Gauge offer over digital gauges?
A: The analog display provides immediate, clear readings without the need for power, and its robust construction offers exceptional durability in demanding industrial conditions, making it ideal for reliable, continuous monitoring.
Q: How should I maintain or clean the gauge window and case?
A: Use a soft cloth with mild detergent to clean the clear acrylic window and powder-coated aluminum case. Avoid abrasive cleaners to preserve clarity and finish.
Q: What is the overpressure tolerance, and how does it protect my system?
A: This gauge can withstand up to 200% of its maximum scale range (0-250 MM WC), offering a substantial margin of safety in the event of transient pressure spikes.